Mallinckrodt NPB-190 Manual do Operador Página 26

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 79
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 25
Setup
Figure 3: Connecting the Power Cord
3. Verify that the supply voltage selection switch located on the
rear of the monitor matches the AC power requirement at
your location.
4. Plug the male connector end of the power cord into a properly
grounded AC outlet.
WARNING: In the USA, do not connect the monitor to an
electrical outlet controlled by a wall switch.
5. Select a Nellcor sensor appropriate for the patient to be
monitored (see the Sensors and Accessories section of this
manual for sensor selection information).
WARNING: Do not use an extension cable with the NPB-290
monitor (other than the MC-10 or SCP-10 sensor cable). Use
of another sensor extension cable will have an adverse effect
on performance. Do not attach to the sensor port any cable
that is intended for computer use.
6. Plug the sensor into the MC-10 or SCP-10 sensor cable, and
secure the sensor in place by lowering the plastic sensor lock
over the sensor connector until it clicks into place (refer to the
MC-10 or SCP-10 directions for use).
18
Vista de página 25
1 2 ... 21 22 23 24 25 26 27 28 29 30 31 ... 78 79

Comentários a estes Manuais

Sem comentários