Start-Up and Use
Note: As the battery is used and recharged over a period of
time, the amount of time between the onset of the low
battery alarm and the instrument shut-off may become
shorter.
If the backlight is turned off during a low battery condition, it
cannot be turned back on.
It is recommended that the internal battery be replaced by
qualified service personnel every 24 months.
Caution: If the NPB-295 is to be stored for a period of 2
months or longer, notify service personnel to remove the
battery from the monitor prior to storage. Recharge the
battery when it has not been charged for 2 or more months.
DISPOSAL OF DEVICE COMPONENTS
Caution: Follow local governing ordinances and recycling
instructions regarding disposal or recycling of device
components, including batteries.